www.stryker.com 3006-209-001 REV B 67
Preventive maintenance
Remove the product from service before you perform the preventive maintenance inspection. Make sure that all items
listed during annual preventive maintenance for all Stryker Medical products. You may need to perform preventive
maintenance checks more often based on your level of product usage. Service only by qualified personnel.
Note: Clean and disinfect the exterior of the support surface before inspection, if applicable.
Inspect the following items:
All fasteners are secure
Engage brake pedal and push on bed to make sure that all casters lock securely
LED on the footboard and head end siderails blink when brakes are not engaged
Locking steer caster locks and unlocks
Siderails move, latch, and stow
CPR release operable
Foot prop intact and operable
IV pole is intact and operable (optional equipment)
Foley bag hooks intact
No cracks or splits in headboard, footboard or siderail panels
No rips or cracks in mattress cover
All functions on head end siderails operable (includes LEDs)
All functions on footboard operable (includes LEDs)
Scale and bed exit system calibrated
Motion interrupt switches operable
Night light operable
Power cord and plug not frayed or damaged
No cables worn or pinched
All electrical connections tight
All grounds secure to the frame
Ground impedance not more than 200 (milliohms)
Current leakage not more than 300 µA (microamps)
Apply grease to the litter grease points (see maintenance manual for locations)
Motorized drive wheel functions (Zoom option)
Motion release switches operable (Zoom option)
All functions on head end control panel operable (Zoom / battery backup option)
Confirm battery power functions (Zoom / battery backup option)
Make sure of power cord slack at the inlet filter (Zoom / battery backup option)
Make sure that ground chains are clean, intact, and have at least two links touching the floor
Make sure that Fowler angle accuracy is - 60°
Make sure that the Fowler holds position at 30° with patient weight
Siderail switches operable (iBed Awareness option)
Center light bar LED and side light LED operable (iBed Awareness option)
Inspect footboard control labels for signs of degradation
Inspect siderail gas spring for oil leaks
Inspect Fowler damper for oil leaks
Make sure that all motions function
Make sure that the nurse call functions
Make sure that the nurse call battery functions (optional equipment)
Make sure that the iBed Wireless Module and IR Module are intact and footboard icons display (iBed
Awareness option)

Table of Contents

Other manuals for Stryker S3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker S3 and is the answer not in the manual?

Stryker S3 Specifications

General IconGeneral
Weight Capacity500 lbs (227 kg)
Maximum Patient Weight500 lbs (227 kg)
Bed Frame MaterialSteel
TypeHospital Bed
Bed Width35 inches (89 cm)
Caster Size5 inches (12.7 cm)
Electrical Requirements120 VAC, 60 Hz
Width35 inches (89 cm)
Power Requirements120 VAC, 60 Hz
Power BackupBattery backup available
AccessoriesIV pole
SiderailsYes