www.stryker.com 3006-209-001 REV B 31
Operation
Activating nurse call communication (option) (Continued)
Note: A 9V battery, located in the head end of the product, powers the nurse call signal. If the product is unplugged
from a wall outlet while the nurse call pendant port switch is set to the on position (Figure 20 on page 31), the charge of
the 9V battery will begin to drain.
If the 9V nurse call battery needs to be replaced, a message will appear on the footboard display. See Replacing the
nurse call backup battery (option) on page 31.
Figure 20: Nurse call switch
Replacing the nurse call backup battery (option)
To prevent a low battery condition when the bed is not plugged in, position the cord out switch at the head end of the
bed to the off position. If the switch is not positioned as shown below (A) (Figure 21 on page 32) and the bed power cord
and pendant cord are unplugged, the life of the backup battery will be significantly reduced.

Table of Contents

Other manuals for Stryker S3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker S3 and is the answer not in the manual?

Stryker S3 Specifications

General IconGeneral
Weight Capacity500 lbs (227 kg)
Maximum Patient Weight500 lbs (227 kg)
Bed Frame MaterialSteel
TypeHospital Bed
Bed Width35 inches (89 cm)
Caster Size5 inches (12.7 cm)
Electrical Requirements120 VAC, 60 Hz
Width35 inches (89 cm)
Power Requirements120 VAC, 60 Hz
Power BackupBattery backup available
AccessoriesIV pole
SiderailsYes