www.stryker.com 3006-209-001 REV B 41
Operation
Arming or disarming Chaperone bed exit (option) (Continued)
• sound alarm is triggered
• bed exit status alert is displayed on the display screen
To disarm bed exit, press and hold the Arm/Disarm button again (A). See Footboard control panel - Chaperone bed exit
(option) on page 40.
Footboard control panel - Chaperone bed exit with zone control (option)
WARNING
• Do not use bed exit to replace patient monitoring protocol.
• Bed exit is intended only to aid in the detection of a patient exiting the product.
• Do not use bed exit with patients who weigh less than 50 lb (23 kg).
• Power save mode activates after one hour on battery power with no motion release switch activation. Bed exit,
scale, and product motion stops operating when the product enters the power save mode.
A
Zone 1
Allows the patient to move freely on the bed. Alarms when the patient
moves 50 percent of their body weight out of the designated zone.
B Zone 2
Allows for limited movement. Alarms when the patient approaches the
siderail or the foot end of the bed.
C Zone 3
Allows minimal movement. Alarms when the patient moves out of the
tightly restricted zone.
D
Arm/Disarm
Arms or disarms bed exit
E Zone
Changes the zone
Note: The Bed Exit footboard LED indicator illuminates when you arm Chaperone Bed Exit (Figure 27 on page 43).