www.stryker.com 3006-209-001 REV B 45
Operation
Footboard control panel - Scale (Continued)
To weigh a patient:
1. Press and hold the Scale button (A) (see Footboard control panel - Scale on page 44) until Release Button
appears (Figure 28 on page 45).
Figure 28: Release button
2. Release the Scale button.
Note: Do not touch the product when you weigh the patient (Figure 29 on page 45).
Figure 29: Do not touch bed
A confirmation notification indicates that weighing the patient was successful (Figure 30 on page 45).
Figure 30: Patient weight
Note: To clear the patient weight from the display, press the Scale button again. The patient weight is still recorded in
the weight log.
Notes
Always calibrate the bed after adding a support surface or mattress to the bed frame.
Always set the scale to zero before putting a patient on the product.
Setting the scale to zero
The zero function clears all of the stored values from the weight log, change patient weight, and gain or loss.
Note: Always set the scale to zero before putting a patient on the product.