Warnings and Cautions
Please read this manual and follow its instructions carefullyThe words warning, caution,
and note carry special meanings and should be carefully reviewed:
Warning: Indicates measures to avoid potential serious injury to the user and the
patient and/or damage to this device
Caution: Indicates risks to the equipment� Failure to follow cautions may result in
product damage�
Note: Provides special information to clarify instructions or present additional useful
information�
1� Caution: Federal law (United States of America) restricts this device to use by, or on order of,
a physician�
2� Read this operating manual thoroughly and be familiar with its contents prior to setting up
or using this equipment�
3� Carefully unpack the unit and check for any damage that may have occurred during
shipment� If damage is detected, contact your Stryker representative�
4� Use of this device is restricted to qualied health care professionals who have a complete
knowledge of the use of this equipment and the devices to which it can be connected�
5� Test this equipment prior to using it in a surgical procedure� This unit was fully tested at the
factory before shipment; however, it must be tested for proper function in the environment
and setup conguration in which it will be used
6� To avoid risk of electric shock, use only a hospital-grade power cord furnished with the
unit� Disconnect the unit from the power supply to make connections or to inspect the
equipment�
7� To avoid the risk of electric shock, this equipment must only be connected to a supply mains
with protective earth�
8� Do not position the console so that it is dicult to disconnect the power cord from the
supply mains�
9� Install this device in an operating room that complies with all applicable IEC, CEC, and NEC
requirements for safety of electrical devices� Any installation or connection with other
devices shall be evaluated for electrical safety according to the IEC 60601-1 standard�
10� Use only clean, dry, undamaged Stryker brand media� The use of damaged, wet, or
incompatible media will void the warranty and may result in the malfunctioning of this
equipment�
11� Make sure that all equipment connected to the SDC3 is properly grounded (and isolated, if
required)�
12� To avoid risk of re, do not use this device in the presence of ammable anesthetics
13� Avoid direct exposure to the laser beam of the DVD/CD drive as it may harm the eyes
14� Pay close attention to the care and cleaning instructions in this manual� Failure to follow
these instructions may result in product damage�
15� Do not modify this equipment without authorization of the manufacturer

Table of Contents

Other manuals for Stryker SDC3

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Stryker SDC3 and is the answer not in the manual?

Stryker SDC3 Specifications

General IconGeneral
BrandStryker
ModelSDC3
CategoryMedical Equipment
LanguageEnglish

Related product manuals