EasyManuals Logo

Tektronix Keithley 4225-PMU User Manual

Tektronix Keithley 4225-PMU
187 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #75 background imageLoading...
Page #75 background image
Section
3: Setting up PMUs and PGUs in Clarius Model 4200A-SCS Pulse Card (PGU and PMU)
User's Manual
3-34 4200A-PMU-900-01 Rev. B March 2023
Select Shift+click and move or scroll a magnified graph by moving the mouse.
Figure 63: Scroll (or move) a magnified entire test graph
Errors
If either graph is not updating as expected, select Refresh to zoom out and redraw both graphs. It is
possible to have a test with too many pulses to be suitably graphed. This may be from too many
pulses from a large number of sweep points or step points, or a large number of pulses. The following
figure shows the output “Too many pulses to graph” for the lower graph. This does not mean that the
test will not run. If you select OK on the PMU Advanced Test Settings dialog and it does not generate
any warning or error messages, the test is valid and runs (even though the entire test waveform
cannot be displayed).
Figure 64: Preview error (Too many pulses to graph)

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix Keithley 4225-PMU and is the answer not in the manual?

Tektronix Keithley 4225-PMU Specifications

General IconGeneral
BrandTektronix
ModelKeithley 4225-PMU
CategoryComputer Hardware
LanguageEnglish