Commands
TDS5000B Series Service Manual
5-11
If the current step is the last step and the calibration succeeds, new calibration
data and pass status is saved to non-volatile memory and the oscilloscope returns
to normal operation in the same state it was in before beginning factory
calibration.
If the current step is the last step and any factory calibration step has failed,
without a successful pass for repeated steps, a fail status will be saved to
non-volatile memory. Factory calibration data will not be saved to non-volatile
memory if the factory status is FAIL.
CALibrate:FACtory:STARt
CALibrate:FACtory:CONTinue
CALIBRATE:FACTORY:CONTINUE
causes the instrument to execute the next step in the calibration procedure.
CALibrate:FACtory:FANcontrol
This command sets or returns the status of the fan control.
CALibrate:FACtory:FANcontrol <NR3> [, <NR3>]
<NR3> sets the temperature in degrees Celsius
<NR3> sets the mean sea level specified in feet; this command is optional and if
not specified, it defaults to 0.0 (sea level)
CALibrate:FACtory:FANcontrol?
0 means that the fan was never calibrated
1 means that calibration is in process
2 means that the fan has been calibrated
CALibrate:FACtory:FANcontrol 20
This sets the fan control calibration temperature to 20 degrees Celsius at sea
level.
CALibrate:FACtory:FANcontrol?
Related C ommands
Syntax
Example
Syntax 1
Arguments
Syntax 2
Returns
Example 1
Example 2