Commands
5-12
TDS5000B Series Service Manual
This query might return
:CALIBRATE:FACTORY:FANCONTROL 1
indicating that calibration is in process.
CALibrate:FACtory:NOTify:DUE?
This command returns whether the factory calibration is due.
CALibrate:FACtory:NOTify:DUE?
0 <NR1> means that the oscilloscope is not due for factory calibration.
1 <NR1> means that the oscilloscope is due for factory calibration.
CALIBRATE:NOTIFY:DUE?
might return
:CALIBRATE:FACTORY:NOTIFY:DUE 0
signifying that factory calibration is not due.
CALibrate:FACtory:NOTify:HOURs
This command sets or returns the number of hours until the user is notified that
the instrument is due for factory calibration.
CALibrate:FACtory:NOTify:HOURs {INFInite|<NR1>}
INFInite means whatever the specified maximum is for the instrument.
<NR1> set to an integer value.
CALIBRATE:NOTIFY:HOURS 1500
sets the calibration notification time to 1,500 hours.
CALIBRATE:FACTORY:NOTIFY:HOURS?
might return
Syntax
Returns
Example
Syntax
Argument
Example 1
Example 2