EasyManuals Logo

Tektronix TDS5000B Series Service Manual

Tektronix TDS5000B Series
180 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #49 background imageLoading...
Page #49 background image
Commands
TDS5000B Series Service Manual
5-13
:CALIBRATE:FACTORY:NOTIFY:HOURS 2000
signifying that the calibration due notifier will be displayed 2000 hours after the
last calibration.
CALibrate:FACtory:NOTify:YEARs
This command sets or returns the number of years until the user is notified that
the instrument is due for factory calibration.
CALibrate:FACtory:NOTify:YEARs {INFInite|<NR3>}
INFInite means whatever the specified maximum is for the instrument.
<NR1> set to a floating point value.
CALIBRATE:FACTORY:NOTIFY:HOURS 1.5e0
sets the calibration notification time to 1.5 years.
CALIBRATE:FACTORY:NOTIFY:HOURS?
might return
:CALIBRATE:FACTORY:NOTIFY:HOURS 1.5e0
signifying that the calibration due notifier will be displayed 1.5 years after the
last calibration.
CALibrate:FACtory:PREVious
This command backs up the calibration steps one step so that the next C ALi-
brate:FACtory:CONTInue command will repeat the factory calibration step that
just finished. Repeated CALibrate:FACtory:PREVious commands can back up as
many calibration steps as desired, at least until step one is reached. The
appropriate notifier, with signal source information, is displayed after the factory
calibration step is adjusted.
CALibrate:FACtory:PREVious
CALIBRATE:FACTORY:PREVIOUS
backs up the calibration steps one step.
Syntax
Argument
Example 1
Example 2
Syntax
Example

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals