Commands
5-14
TDS5000B Series Service Manual
CALibrate:FACtory:STARt
This command starts the factory calibration. The factory calibration process
consists of a series of steps. Do not send any other commands that change the
state of the oscilloscope until the calibration process is complete.
CALibrate:FACtory:ABOrt
CALibrate:FACtory:STARt
CALIBRATE:FACTORY:START
starts the factory calibration.
CALibrate:FACtory:STATus?
This command returns the factory calibration status.
CALibrate:FACtory:STATus?
FAIL if the factory calibration failed
PASS if the factory calibration passed
UNCAL if the factory calibration has not been run
CALIBRATE:FACTORY:STATUS
might return
:CALIBRATE:FACTORY:STATUS PASS
indicating the factory calibration passed
CALibrate:FACtory:STEPSTATus
This command returns the status of the current calibration step.
It returns pass during oscilloscope power on without regard to actual calibration
status. It returns pass or fail for the factory calibration step that most recently
executed when factory calibration is in progress. Use this query to synchronize
programmed factory calibration steps.
Related C ommands
Syntax
Example
Syntax
Returns
Example

Table of Contents

Other manuals for Tektronix TDS5000B Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Tektronix TDS5000B Series and is the answer not in the manual?

Tektronix TDS5000B Series Specifications

General IconGeneral
BrandTektronix
ModelTDS5000B Series
CategoryTest Equipment
LanguageEnglish

Related product manuals