Commands
TDS5000B Series Service Manual
5-21
DIAg:CONTROL:LOOP
This command (no query form) determines whether the next execution of
diagnostics executes once or continuously loops on the selected set of diagnos-
tics (assuming the halt control is set to off using the DIAg:CONTROL:HALT
command or that the halt control is set to ON but no failures occur). This
command is equivalent to selecting Instrument Diagnostics from the Utilities
menu and then enabling Loop Control.
DIAg:CONTROL:HALT, DIAg:STOP
DIAg:CONTROL:LOOP (ON|OFF|<NR1>)
ON enables the loop function, causing the execution of diagnostics to continu-
ously loop.
OFF disables the loop function, causing the oscilloscope to execute the entire set
of diagnostics once and then halt.
<NR1>--A 0 enables the loop function; any other number disables the loop
function.
DIAg:CONTROL:LOOP ON
This command enables the loop function.
DIAg:EXECUTE
This command (no query form) starts the execution of the currently selected set
of diagnostics. This command is equivalent to selecting Instrument Diagnostics
from the Utilities menu and then pressing Run.
DIAg:STATE
DIAg:EXECUTE
DIAg:EXECUTE
This command starts the execution of the selected set of diagnostics.
Related C ommands
Syntax
Arguments
Examples
Related C ommands
Syntax
Example