Commands
5-32
TDS5000B Series Service Manual
NOTE. This command should be executed only if DIAg:LEVEL is currently set to
SUBSYS.
DIAg:SELect:AREA, DIAg:SELect:TEST, DIAg:SELect:LAS T
DIAg:SELect:SUBSYS <NR1>
<NR1> selects a subsystem by number, which can range from 0 (zero selects
ALL) through 15 (as specified by Diag:NUMITEMS).
DIAg:SELect:SUBSYS
selects Subsystem 1 as the subsystem to be diagnosed.
DIAg:SELect:TEST
This command (no query form) selects one of the available tests. This command
is equivalent to selecting Instrument Diagnostics from the Utilities menu and
then choosing a Test from the drop down list.
NOTE. This command should be executed only if DIAg:LEVEL is currently set to
TEST.
DIAg:SELect:AREA, DIAg:SELect:SUBSYS, DIAg:SELect:LAST
DIAg:SELect:TEST <NR1>
<NR1> selects a test by number , which can range from 0 (zero selects ALL)
through 15 (as specified by DIAg:NUMITEMS).
DIAg:SELect:TEST 1
selects Test 1 as the test to be executed.
Related C ommands
Syntax
Arguments
Example
Related C ommands
Syntax
Arguments
Example