Commands
TDS5000B Series Service Manual
5-33
DIAg:STATE
This command (no query form) changes the oscilloscope operating state.
Depending on the argument, diagnostics capabilities are either turned on or off.
This command is equivalent to opening the DIAg:STATE dialog (ON) or closing
it (OFF).
TEST:S TOP
DIAg:STATE {ON|OFF}
ON puts the instrument into the state in which diagnostics can be run. This
argument is used automatically if either DIAg:EXECUTE or DIAg:TEST
commands are executed.
OFF disables diagnostics capabilities and returns the oscilloscope to a normal
operating state.
DIAg:STATE OFF
turns off diagnostics capabilities and returns the instrument to a normal operating
state.
DIAg:STOP
This command (no query form) causes diagnostics (or test) execution to
terminate at the end of the next low-level test. This command is equivalent to
selecting Instrument Diagnostics from the Utilities menu and then clicking
Abort.
TEST:S TOP
DIAg:STOP
DIAg:STOP
terminates diagnostics (or test) execution at the end of the next low-level test.
Related C ommands
Syntax
Arguments
Example
Related C ommands
Syntax
Example