142 Maintenance and Service 083730300A DCN8101
Can I use the IZS option to
calibrate the analyzer?
Yes. However, whereas this may be acceptable for basic calibration
checks, the IZS option is not as accurate as the external span and zero
gas bottles.
To achieve highest accuracy, it is recommended to use cylinders of
calibrated span gases in combination with a zero air source.
Q: What is the averaging time for
an T300/T300M?
A: The default averaging time, optimized for ambient pollution
monitoring, is 150 seconds for stable concentrations and 10 seconds for
rapidly changing concentrations (see Section 6.5.1 for more
information).
However, it is adjustable over a range of 0.5 second to 200 seconds
(please contact Technical Support for more information).