083730300A DCN8101 Communications and Remote Operation 89
Table 3-3. Default Hessen Status Flag Assignments
STATUS FLAG NAME DEFAULT BIT ASSIGNMENT
WARNING FLAGS
SAMPLE FLOW WARNING
0001
BENCH TEMP WARNING
0002
0004
BOX TEMP WARNING
0008
WHEEL TEMP WARNING
0010
SAMPLE TEMP WARN
0020
0040
INVALID CONC
(The Instrument’s Front Panel Display Will Show The
Concentration As “Warnings”)
0080
OPERATIONAL FLAGS
1
Instrument OFF
0100
In MANUAL Calibration Mode
0200
In ZERO Calibration Mode
4
0400
In O
2
Calibration Mode (if O
2
sensor installed )
2,4
0400
In CO
2
Calibration Mode (if CO
2
sensor installed )
2,4
0400
In SPAN Calibration Mode
0800
UNITS OF MEASURE FLAGS
0000
2000
PPB
4000
6000
SPARE/UNUSED BITS
1000, 8000
AZERO WARN
2
CANNOT DYN SPAN
2
CANNOT DYN ZERO
3
1
CONC ALARM 1
3
SYSTEM RESET
1
CONC ALARM 2
3
1
These status flags are standard for all instruments and should probably not be modified.
2
Only applicable if the optional internal span gas generator is installed.
3
Only applicable if the analyzer is equipped with an alarm options.
3
It is possible to assign more than one flag to the same Hessen status bit. This allows the grouping of
similar flags, such as all temperature warnings, under the same status bit.
Be careful not to assign conflicting flags to the same bit as each status bit will be triggered if any of the
assigned flags is active.