4.Methods of Connection
27
CODE1 ~ 16:
Specifies operation ch for Hold mode, High/Low Limit, Comparison Waveform, etc.
(See "10-2. Setting and Operating Method" P47, "11-2. Setting and Operating
Method" P64, "12-2. Setting and Operating Method" P65, and "13-2. Setting and
Operating Method" P68.)
STROBE:
Changing of operation ch by changing CODE 1 ~ 16 is prohibited when this signal is
ON. (Previous operation ch is maintained.) Use to prevent chattering when changing
operation ch.
Select output:
Output pin 44 ~ 46 (LO/OK/HI displacement 2) operate when this signal is ON.
- Pin 44/45 "Status" output
For logic, see "4-3-4. Equivalent Circuit (Output)" P30. )
- Pin 46 "Level Edge" output (Present when this signal is ON. Use for level
edge clear, etc.)
When this signal is OFF, output pin 44 ~ 46 result in the operation of LO/OK/HI
displacement 2.
Judgment output signals
The results of High/Low Limit Judgment in each Operation Mode are output.
LO1, OK1, HI1 1) When the Multi-Hold / Hysteresis mode is used.
Outputs the results of judgment of the load with respect to the
high limit / low limit.
2)When the Waveform Comparison / Waveform and Displace-
ment Comparison mode is used.
Outputs the results of judgment of the load with respect to the
high and low limit waveforms.
3) When the Hysteresis 2 mode is used.
Outputs the results of judgment of the load with respect to the
high limit / low limit, and the results of judgment of the
difference between go and return with respect to the hi-hi
limit / lo-lo limit.
Status Pin 44 Pin 45
WAITS / STOP 0 0
WAI T2 1 0
WA IT L 0 1
SAMP / COPY 1 1