10.Multi-Hold Mode
47
10. Multi-Hold Mode
10-1. Function
In the Multi-Hold mode, necessary points are detected in the displayed waveform to
make a judgment such as a High/Low Limit Comparison. Up to 32 channels of hold
types, High and Low Limits, etc. can be stored, which can be switched by external
signals or can automatically be switched by using the Auto Code Up function. Detected
hold points are displayed on a graph with a "+" mark.
10-2. Setting and Operating Method
1. Set desired value(s) [0 ~ 31 (32: all channel number selection)] under Channel
Number Selection in Code Setting, and set each item such as the High and Low
Limits.
2. Input the channel number(s) set in "1." by using channel selection signals (CODE 1
~ 16) to set operation channel(s).
3. Sampling can be performed with the value(s) set in "1.".
4. By also inputting set value(s) to other channel number(s) in "1." beforehand,
sampling can be performed under various set conditions by switching operation
channels (CODE 1 ~ 16) during sampling.
Example
0123
Peak Hold
Inflection A Valley HoldSample