Calibrating
the
Force
FX-C
Generator
5-28
Step
3 -
Adjust
the
Clock
The
hour
and
minute
values
are
stored
in
the
real-time clock
on
the
Control
board.
The
clock
was
originally
set
for
Mountain
Standard
Time.
The
clock
is
configured to the 24-hour (i.e., military time) format.
A.
Verify
that
the
Bipolar
display
shows
calibration
step
number
3.
B.
To
select the hour,
press
the
Coag
up
(
~
)
or
down
(
V'
)
arrow
buttons
until
the
correct
value
(0-23)
appears
in
the
Coag
display.
Examples:
For 2 AM,
set
the
hour
value
to 2.
For
2 PM,
set
the
value
to 14.
C.
To
display
the
minute
value,
press
the
Med
(Fulgurate)
button.
D.
To
select
the
minute
value,
press
the
Coag
up
(
~)
or
down
(
V'
)
arrow
buttons
until
the
correct
value
(0-59)
appears
in
the
Coag
display.
E.
To
save
the
hour
and
minute
values
and
go
to
the
next
calibration step,
press
the
Bipolar
up
arrow
(~)
button.
Next
check the REM
impedance
as
described
in
Step 4 - Check and
Adjust
the
REM
Oscillator Frequency and Impedance.