Periodic
Safety
Check
"'Important
If
the
coag
mode
has
been
optionally
changed
to
default
to
Desiccate
or
Spray,
that
corresponding
indicator
illuminates
after
the
self-test
is
performed
successfully
5-16
Tes'ting
the
Generator
Turning
on
the
generator
initiates
an
internal
self-test to verify
the
calibration.
The
self-test also checks
the
operation
of
the
speaker, all
indicators,
and
the
displays.
Warning
l'{J$eth~
'Otherwi$
1. Turn
on
the
generator
by
pressing
the
front
panel
On
( I ) switch. Verify
the
following:
All
visual
indicators
and
displays
on
the
front
panel
illuminate.
Activation tones
sound
to verify
that
the
speaker
is
working
properly.
2.
If
the self-test is successful, a tone
sounds.
Verify
the
following:
Indicators
above
the
default
mode
buttons
(Standard bipolar,
Pure
cut,
and
Fulgurate
coag) illuminate green.
Each display
shows
a
power
setting
of
one
watt.
The
REM
Alarm
indicator illuminates red.
If
the self-test is not successful,
an
alarm
tone
sounds.
A
number
may
momentarily
appear
in
the
Cut
display
and,
in
most
cases,
the
generator is disabled.
Note
the
number
and
refer to Responding to
System
Alanns
in
Section
6.
If
you
removed
and/or
replaced
the
battery,
alarm
number
212
may
appear
in
the
Cut
display
when
you
turn
on
the
generator.
If
this
happens,
calibrate
the
generator.
Force
f)('M-C
Service
Manual