Valleylab Force FX-C - Standards and IEC Classifications; Class I Equipment(IEC 601-1); Type CF Equipment (IEC 601-1);Defibriliator Proof; Drip Proof (IEC 601-2-2)

Valleylab Force FX-C
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Standards
and
IEC
Classifications
Standards
and
IEC
Classifications
Force
FXTM-C
Service
Manual
Symbol
Caution
Indicates
ATTENTION
Consult
accompanying
documents.
The
generator
output
is
floating
(isolated)
with
respect
to
ground.
DANGER
Explosion
risk
if
used
with
flammable
anesthetics.
To
reduce
the
risk
of
electric
shock,
do
not
remove
the
cover.
Refer
servicing
to
qualified
service
personnel.
Class
I
Equipment
(IEC
601-1)
Accessible
conductive
parts
cannot
become
live
in
the
event
of
a
basic
insulation
failure
because
of
the
way
in
which
they
are
connected
to
the
protective
earth
conductor.
Type
CF
Equipment
(IEC
601-1)/Defibrillator
Proof
The
Force FX-C
generator
provides
a
high
degree
of
protection
against
electric
shock,
particularly
regarding
allowable
leakage
currents.
It
is
type
CF
isolated
(floating)
output
and
may
be
used
for
procedures
involving
the
heart.
The
Force FX-C
generator
patient
return
electrode
terminal
is
protected
from
defibrillator
discharge
according
to
ANSI/
AAMI
HFl8
and
IEC 601-
2-2.
Drip
Proof
(IEC
601-2-2)
The
generator
enclosure
is
constnlcted
so
that
liquid
spillage
in
normal
use
does
not
wet
electrical
insulation
or
other
components
which,
when
wet,
are
likely
to
affect
adversely
the
safety
of
the
generator.
3-7

Table of Contents

Related product manuals