Valleylab Force FX-C - Section 6. Troubleshooting; Inspecting the Generator

Valleylab Force FX-C
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Troubleshooting
If
the
generator
is
not
functioning
properly,
use
the
information
in
this
section
to
perform
the
following
tasks:
Identify
and
correct
the
malfunction.
If a
system
alarm
number
was
displayed,
take
the
appropriate
action
to
correct
the
alarm
condition.
Inspecting
the
Generator
Force
FXTM-C
Service
Manual
If
the
Force
FX-C
generator
malfunctions,
check
for
obvious
conditions
that
may
have
caused
the
problem:
Check
the
generator
for
visible
sign
..
s
of
physical
damage.
Verify
that
all
accessory
cords
are
properly
connected.
Check
the
power
cord.
Replace
the
power
cord
if
you
find
exposed
wires,
cracks,
frayed
insulation,
or
a
danlaged
connector.
Open
the
fuse
drawer
and
inspect
the
fuse
housing
and
fuses
for
damage
and
corrosion.
Verify
that
the
fuses
are
firmly
seated
An
internal
component
malfunction
in
the
generator
can
damage
the
fuses. You
may
need
to
replace
the
fuses
if
the
generator
fails
the
self-test
or
stops
functioning.
Refer
to
Fu.''f Replacement
in
Section
7.
6-1

Table of Contents