Performance
Characteristics
3-6
High
Frequency
(RF)
Leakage
Current
Monopolar
RF
leakage
current
(additional
tolerance)
Input
Power
110-120
Volt
<
150
mArms
Maximum
VA
at
nominal
line
voltage:
Idle:
52
VA
Bipolar:
450
VA
Cut:
924
VA
Coag:
530
VA
Input
mains
voltage,
full
regulation
range:
104-132
Vac
Input
mains
voltage,
operating
range:
85-132
Vac
Mains
current
(maximum):
Idle:
0.4
A
Bipolar:
2,0
A
Cut:
7.0
A
Coag:
4.0A
Mains
line
frequency
range
(nominal):
50
to
60
Hz
Fuses
(2):
F8
A
Power
cord:
3-prang
hospital
grade
connector
220-240
Volt
Maximum
VA
at
nominal
line
voltage:
Idle:
52
VA
Bipolar:
450
VA
Cut:
924
VA
Coag:
530VA
Input
mains
voltage,
full
regulation
range:
208-264
Vac
Input
mains
voltage,
operating
range:
170-264
Vac
Mains
current
(maximum):
Idle:
0.2
A
Bipolar:
1.0
A
Cut:
3.5
A
Coag:
2.0A