Valleylab Force FX-C - Check the Output for the Coag Modes

Valleylab Force FX-C
214 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Periodic
Safety
Check
5-20
Step
2 -
Check
the
Output
for
the
Coag
Modes
A.
Complete
items
A
and
B
of
the
previous
step, replacing 300
ohm
resistor
,vith
500
ohm
resistor
in
item
B.4.
B.
Press
the
Low
(Desiccate) button.
C.
Press
the
Coag
up
(
L\
)
or
down
( V )
arrow
buttons
to
set
the
coag
power
to 80
watts.
D.
Test
the
monopolar
coag
output.
(1)
Press the footswitch coag
pedal
and,
while
activating
the
generator,
note
the
output
on
the
voltmeter.
(2)
Release
the
footswitch pedal.
(3)
Based
on
the
voltmeter
setting
and
the
current
transformer
you
are
using, calculate
and
record
the
output
current.
E.
Press
the
Med
(Fulgurate)
button
and
repeat
step
2.D.
F.
Press
the
Med
(Fulgurate)
button
and
hold
for
two
seconds.
After
the
tone
sounds
and
the
"L"
appears
on
the
left
side
of
the
Coag
display,
repeat
step
2.D.
G.
Press the
High
(Spray)
button
and
repeat
step
2.D.
H.
Verify
that
the
generator
output
for
each
mode
is
400 ± 28
rnA
rms.
If
the
output
is
outside
the
specified range, calibrate
the
monopolar
output
as
described
in
calibration steps 5,
6,
7,
and
8.
Then
repeat
this
procedure.
If
the
output
for
one
or
more
coag
modes
remains
outside
the
specified range, call
the
Valleylab Service Center.
Force
F)(TM-C
Service
Manual

Table of Contents

Related product manuals