Number
260
261
262
Description
Internal
diagnostics.
AJD
conversion
did
not
complete
in
allowed
time.
Software
malfunction.
Delay
time
out
of
bounds
on
main
microcrocontrolier.
Software
malfunction.
Delay
time
out
of
bounds
on
feedback
microcontroller.
270
Software
malfunction.
Data
type
in
me_rtc_gecO
outside
of
the
state
range.
271
Software
malfunction.
Data
type
in
mertc.puU}
outside
of
the
state
range.
451
The
Internal
temperature
limit
was
exceeded
due
to
length
of
activation
time.
Force
FXTM-C
Service
Manual
Responding
to
System
Alarms
Recommended
Action
1.
Replace
the
Control
board.
Refer
to
Control
Board
Replacement
in
Section
7
for
instructions.
2.
Calibrate
the
generator.
Refer
to
Section
5
for
instructions.
If
the
alarm
number
reappears,
record
the
number
and
call
the
VaUeylab
Service
Center.
Turn
off,
then
turn
on
the
generator.
If
the
alarm
number
reappears,
record
the
number
and
call
the
Valleylab
Service
Center.
Verify
that
the
location
of
the
generator
allows
for
adequate
cooling
Use
the
lowest
power
setting
that
achieves
the
desired
effect.
Limit
activation
times,
if
possible.
6-19