Important
If
the
coag
mode
has
been
optionally
changed
to
default
to
Desiccate
or
Spray
that
corresponding
indicator
illuminates
after
the
self-test
is
performed
successfully
Force
FXTM-C
Service
Manual
Setting
Up
the
Generator
7.
If
the
self-test
is
successful,
a
tone
sounds.
Verify
the
following:
Indicators
above
the
default
mode
buttons
(Standard
bipolar,
Pure
cut,
and
Fulgurate
coag)
illuminate
green.
Each
display
shows
a
power
setting
of
1
watt.
The
REM
Alarm
indicator
illuminates
red.
If
the
self-test is
not
successfuL
an
alarm
tone
sounds.
A
number
may
momentarily
appear
in
the
Cut
display
and,
in
most
cases,
the
generator
is
disabled.
Note
the
number
and
refer
to
Reo:ponding
to
System Alarms
in
Section
6.