18-1
IM 765501-01E
Specifications
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
18
17
16
Index
App
18.1 Source Section
DC Voltage Source
Range Setting Source Range Resolution Max. Load Accuracy (One Year) Temperature Coefficient
Current ±(% of setting + V + V×Io/ ±(% of setting + V + V×Io/
f.s of the limiter range) f.s of the limiter range)/°C
200 mV ±205.000 mV 1 µV ±3.2 A 0.02+200 µV+80 µV (400 µV) 0.002+20 µV+8 µV (40 µV)
2 V ±2.05000 V 10 µV ±3.2 A 0.02+300 µV+100 µV (500 µV) 0.002+30 µV+10 µV (50 µV)
12 V ±12.0000 V 100 µV ±3.2 A 0.02+2 mV+800 µV (3 mV) 0.002+200 µV+80 µV (300 µV)
20 V ±20.5000 V 100 µV ±2 A 0.02+2 mV+800 µV (5 mV) 0.002+200 µV+80 µV (500 µV)
30 V ±30.000 V 1 mV ±2 A 0.02+20 mV+5 mV (30 mV) 0.002+2 mV+500 µV (3 mV)
60 V ±60.000 V 1 mV ±1 A 0.02+20 mV+6 mV (40 mV) 0.002+2 mV+600 µV (4 mV)
110 V ±110.000 V 1 mV ±0.5 A 0.02+20 mV+8 mV (70 mV) 0.002+2 mV+800 µV (7 mV)
Accuracy: One year accuracy at 23±5°C.
Temperature coefficient: Add the temperature coefficient at 5 to 18°C and 28 to 40°C.
The values inside the parentheses are those when the limiter range is 3 A.
DC Current Source
Range Source Resolution Max. Load Accuracy (One Year) Temperature
Setting Range Voltage ±(% of setting+A) Coefficient
±(% of setting+A)/°C
20 µA ±20.5000 µA 100 pA ±110 V 0.03+50 nA 0.003+5 nA
200 µA ±205.000 µA1 nA ±110 V 0.03+300 nA 0.003+30 nA
2 mA ±2.05000 mA 10 nA ±110 V 0.03+3 µA 0.003+300 nA
20 mA ±20.5000 mA 100 nA ±110 V 0.03+30 µA 0.003+3 µA
200 mA ±205.000 mA 1 µA ±110 V 0.03+300 µA 0.003+30 µA
0.5 A ±0.50000 A 10 µA ±110 V 0.03+5 mA 0.003+500 µA
1 A ±1.00000 A 10 µA ±60 V 0.03+5 mA 0.003+500 µA
2 A ±2.00000 A 10 µA ±30 V 0.03+5 mA 0.003+500 µA
3 A ±3.20000 A 10 µA ±12 V 0.03+5 mA 0.003+500 µA
Accuracy: One year accuracy at 23±5°C.
Temperature coefficient: Add the temperature coefficient at 5 to 18°C and 28 to 40°C.
Current Limiter
|Limiter Value|
1
Range Setting Resolution Minimum Limit Value
2
0.10 µA to 20.00 µA 20 µA 10 nA 100 nA
20.1 µA to 200.0 µA 200 µA 100 nA 1 µA
0.201 mA to 2.000 mA 2 mA 1 µA 10 µA
2.01 mA to 20.00 mA 20 mA 10 µA 100 µA
20.1 mA to 200.0 mA 200 mA 100 µA1 mA
0.201 A to 3.200 A 3 A 1 mA 10 mA
1 When tracking is OFF, |Hi| or |Lo|, whichever is greater.
2 Minimum value when tracking is OFF.
Voltage Limiter
|Limiter Value|
3
Range Setting Resolution Minimum Limit Value
4
1.0 mV to 200.0 mV 200 mV 100 µV1 mV
0.201 V to 2.000 V 2 V 1 mV 10 mV
2.01 V to 20.00 V 20 V 10 mV 100 mV
20.1 V to 110.0 V 110 V 100 mV 1 V
3 When tracking is OFF, |Hi| or |Lo|, whichever is greater.
4 Minimum value when tracking is OFF.
Chapter 18 Specifications