EasyManuals Logo

YOKOGAWA 765501 User Manual

YOKOGAWA 765501
277 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #187 background imageLoading...
Page #187 background image
16-15
IM 765501-01E
Communication Commands
3
2
1
6
5
4
9
8
7
12
11
10
15
14
13
18
17
16
16
Index
App
16.2.3 Source Commands (SOURce GRoup)
:SOURce:DELay
Function Sets the source delay or queries the current
setting.
Syntax :SOURce:DELay <Time>|MINimum|
MAXimum
<Time> = Sets the source delay to the specified
value.
MINimum = Sets the source delay to the
minimum value.
MAXimum = Sets the source delay to the
maximum value.
:SOURce:DELay? [MINimum|MAXimum]
No parameter = Queries the current value.
MINimum = Queries the minimum value.
MAXimum = Queries the maximum value.
<Time>
Example :SOUR:DEL 2.5E-3
:SOUR:DEL MIN
:SOUR:DEL?
:SOUR:DEL? MAX
:SOURce:PULSe:WIDTh
Function Sets the pulse width for pulse generation or
queries the current setting.
Syntax :SOURce:PULSe:WIDTh <Time>|MINimum|
MAXimum
<Time> = Sets the source delay to the specified
value.
MINimum = Sets the source delay to the
minimum value.
MAXimum = Sets the source delay to the
maximum value.
:SOURce:PULSe:WIDTh? [MINimum|
MAXimum]
No parameter = Queries the current value.
MINimum = Queries the minimum value.
MAXimum = Queries the maximum value.
<Time>
Example :SOUR:PULS:WIDT 0.75
:SOUR:PULS:WIDT MIN
:SOUR:PULS:WIDT?
:SOUR:PULS:WIDT? MAX
:SOURce:FUNCtion
Function Sets the source function (voltage or current) or
queries the current setting.
Syntax :SOURce:FUNCtion VOLTage|CURRent
VOLTage = Sets the source function to voltage.
CURRent = Sets the source function to current.
:SOURce:FUNCtion?
VOLT = Currently set to voltage.
CURR = Currently set to current.
Example :SOUR:FUNC VOLT
:SOUR:FUNC?
Description When the source function is changed, the
output (:OUTPut:STATe) is automatically
turned OFF.
:SOURce:SHAPe
Function Sets the source mode (DC or pulse) or queries
the current setting.
Syntax :SOURce:SHAPe DC|PULSe
DC = Sets the source mode to DC.
PULSe = Sets the source mode to pulse.
:SOURce:SHAPe?
DC = Currently set to DC.
PULS = Currently set to pulse.
Example :SOUR:SHAP PULS
:SOUR:SHAP?
Description This function corresponds to MODE on the front
panel.
:SOURce:MODE
Function Sets the source pattern (fixed level, sweep, or
program sweep) or queries the current setting.
Syntax :SOURce:MODE FIXed|SWEep|LIST
FIXed = Sets the source pattern to constant
level (sweep OFF).
SWEep = Sets the source pattern to sweep
(linear or log sweep).
LIST = Sets the source pattern to program
sweep.
:SOURce:MODE?
FIX = Currently set to constant level (sweep
OFF)
SWE = Currently set to sweep (linear or log
sweep)
LIST = Currently set to program sweep.
Example :SOUR:MODE LIST
:SOUR:MODE?
Description This function corresponds to SWEEP on the
front panel. Specify the linear or log setting of
the sweep mode using the :SOURce:
VOLTage:SWEep:SPACing or :SOURce:
CURRent:SWEep:SPACing command.
16.2 Commands