8-7
IM 765501-01E
Computation
8
7
12
11
10
15
14
13
18
17
16
Index
App
8.4 Comparison Operation
Procedure
MEASURE
STORE
AUTO
RANGE
MENU
MATH
NULL
COMPARE
MEASURE
VALUE
LOCAL
POWER
MISC
TIME
KEY LOCK
SHIFT
V
I
MODE
AUTO
RANGE
LIMIT
VS IS
MENU
SOURCE
OUTPUT CONTROL
SWEEP
START
ZERO
SOURCE
TRIG
MODE
OUTPUT
RANGE
Ω
RECALL
DISPLAY
ESC
• To exit the menu during operation, press
ESC at the upper left of the operation
panel.
• In the procedural explanation below, the
phrase “rotary knob, numeric keys, <,
and >” are used to refer to the operation
of selecting or setting items or entering
values using the rotary knob, numeric
keys, , and keys (BS key and
right arrow key). For details on the
operation using the rotary knob, numeric
keys, , and , see sections 3.8 or
3.9.
Turning the Comparison Operation ON/OFF
1. Press SHIFT+MATH(COMPARE) to display the COMPARE menu.
2. Press the Compare soft key to select On or Off. When the comparison
operation is turned ON, the MATH key illuminates.
Setting the Reference Value
3. Press the Upper soft key to display the Upper setup screen.
4. Use the rotary knob or numeric keys and < and > keys to set the upper value.
5. If you use the numeric keys, press a soft key corresponding to the desired unit
or the Enter soft key to confirm the setting.
6. Press the Lower soft key to display the Lower setup screen.
7. Use the rotary knob or numeric keys and < and > keys to set the lower value.