Allowableinputvoltage: ±10VDC(betweenH-L)continuous
Allowable common-mode voltage:
Channel-to-channel: 30 VAC rms
Between input and earth:250 VAC rms (-B12,-B35) 30 VACrms(-NDI)
However, NDIS connector shell is connected to earth potential
CMRR
*
: For integration time of 16.67 ms or more: 120 dB or more
For integration time of 1.67 ms or more: 80 dB or more
(voltageconversionvalueat50/60Hz±0.1%,bridgevoltageof2V)
NMRR: For integration time of 16.67 ms or more: 40 dB or more (50/60
Hz±0.1%)
For integration time of 1.67 ms: 50/60 Hz is not rejected.
(voltage conversion value with bridge voltage of 2 V)
Insulation resistance
*
:Betweeninputandearth: 20MΩormoreat500VDC
Withstand voltage
*
: Between input and earth: 2300 VAC for one minute
Channel-to-channel: 30 VACrms or less
Power consumption: Approximately 3 W (one module)
Weight: Approximately 0.5 kg
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal cover)
Terminal type: -B12, -B35: Clamp, terminal board is detachable
-NDI: NDIS, detachable on individual channels
Applicable wire size: 0.14 to 1.5 mm
2
(AWG26 to 16)(excluding -NDI)
* Not app
licable in the case of the NDIS terminal.
Effects of Operating Conditions
Thespecicationsbelowapplywhentheintegrationtimeis16.67msormore.
Warm-up time: At least 30 minutes after power-up
Effects of ambient temperature:
The effect received by changes in 10°C increments of the ambient
temperature is:
±(0.1%ofrange)orless
Effectsofpoweructuation:
MeetstheaccuracyspecicationsforACpowersupplyinthe
range 90 to 132 V and 180 to 250 V
Effectsofmagneticeld:
TheuctuationinexternalmagneticeldsforAC(50/60Hz)400
A/mis±2%orrangeorless
Effect of position: Horizontal with the feet at the bottom is the rule.
External Dimensions
Unit: mm
57 150.8
133.857
4.8
-B12, -B35 -NDI
131
131
Ifnotspecified,thetoleranceis±3%.However,incasesoflessthan10mm,the
toleranceis±
0.3 mm.
4.7 4-CH, Medium-Speed Strain Input Module Specifications