4-28
IM MX100-01E
4.12 10-CH, Medium-Speed Digital Output Module
Specifications
Style number: S1
Number of outputs: 10
Contact mode: A contact (SPST)
Update interval: Output per 100 ms (not synchronized to the measurement
interval)
Contact capacity: 250 VDC/0.1 A, 250 VAC/2 A, or 30 VDC/2A (resistance load)
Contact life*: 100,000 times at rated load (typical)
20,000,000 times at no load (typical)
* The contact life varies depending on the load conditions and the
environment in which it is used.
Insulationresistance: 20MΩormore(500VDC)betweenoutputterminalandearth
20MΩormore(500VDC)betweentheoutputterminals
Withstand voltage: 2300 VAC (50/60 Hz) for one minute between the output terminal
and earth
2300 VAC (50/60 Hz) for one minute between output terminals
Maximum common-mode voltage:
250 VACrms (50/60 Hz)
Power consumption: Approximately 2 W (when all relays are turned ON)
Terminal type: Clamp, attached and removed in units of 5 channels
Applicable wire size: 0.08 to 2.5 mm
2
(AWG28 to 12)
External dimensions: Approx. 57 (W) × 131 (H) × 151 (D) mm (including the terminal
cover)
Weight: Approximately 0.5 kg
External Dimensions
Unit: mm
57
131
150.8
Ifnotspecified,thetoleranceis±3%.However,incasesoflessthan10mm,the
toleranceis±0.3mm.