vibrations.
Input
DC voltage
20 mV
60 mV
200 mV
2 V
6 V
20 V
100 V
R
1
S
1
B
1
K
1
E
1
J
1
T
1
L
2
U
N
3
W
4
KPvsAu7Fe
Pt100
5
JPt100
5
Pt100 (high res.)
JPt100 (high res.)
Ni100 SAMA
6
Ni100 DIN
6
Ni120
7
Pt100
5
JPt100
5
Pt100 (high res.)
JPt100 (high res.)
Pt50
5
Cu10 GE
8
Cu10 L&N
8
Cu10 WEED
8
Cu10 BAILEY
8
J263B
Level
Contact input
-20.000 to 20.000 mV
-60.00 to 60.00 mV
-200.00 to 200.00 mV
-2.0000 to 2.0000 V
-6.000 to 6.000 V
-20.000 to 20.000 V
-100.00 to 100.00 V
0.0 to 1760.0°C
0.0 to 1820.0°C
-200.0 to 1370.0°C
-200.0 to 800.0°C
-200.0 to 1100.0°C
-200.0 to 400.0°C
-200.0 to 900.0°C
-200.0 to 400.0°C
0.0 to 1300.0°C
0.0 to 2315.0°C
0.0 to 300.0K
-200.0 to 600.0°C
-200.0 to 550.0°C
-140.00 to 150.00°C
-140.00 to 150.00°C
-200.0 to 250.0°C
-60.0 to 180.0°C
-70.0 to 200.0°C
-200.0 to 250.0°C
-200.0 to 250.0°C
-140.00 to 150.00°C
-140.00 to 150.00°C
-200.0 to 550.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
0.0 to 300.0K
Vth = 2.4 V
Measurement Accuracy
Integration Time: 16.67 ms
or More
Highest
Resolution
(1 Digit)
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 1°C)
Except 0 to 100°C: ±3.7°C, 100 to
300°C: ±1.5°C for R and S;
400 to 600°C: ±2°C, less than
400°C: not guaranteed for B
±(0.05% of rdg + 0.7°C)
Except -200°C to -100°C:
±(0.05% of rdg + 1°C)
±(0.05% of rdg + 0.5°C)
Except -200°C to -100°C:
±(0.05% of rdg + 0.7°C) for J and L
±(0.05% of rdg + 0.7°C)
±(0.05% of rdg + 1°C)
±(0.05% of rdg + 0.7K)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 0.7°C)
±(0.05% of rdg + 0.3K)
Threshold level accuracy: ±0.1 V
ON for 100 Ω or less, OFF for 10 kΩ or more
9
Rated
Measurement
Range
1 µV
10 µV
10 µV
100 µV
1 mV
1 mV
10 mV
0.1°C
0.1K
0.1°C
0.01°C
0.1°C
0.1°C
0.01°C
0.1°C
0.1K
1 R, S, B, K, E, J, T: ANSI, IEC 584, DIN IEC 584, JIS C 1602-1981
2 L: Fe-CuNi, DIN43710/U: Cu-CuNi, DIN 43710
3 N: Nicrosil-Nisil, IEC 584, DIN IEC 584
4 W: W•5%RE-W•26%Re (Hoskins Mfg Co)
5 Pt50: JIS C 1604-1981, JIS C 1606-1986/Pt100: JIS C 1604-1989, JIS C 1606-1989, IEC 751, DIN IEC 751/
JPt100: JIS C 1604-1981, JIS C 1606-1989
6 SAMA/DIN
7 McGRAW EDISON COMPANY
8 Guaranteed accuracy range Cu10 GE: -84.4 to 170.0°C/Cu10 L&N: -75.0 to 150.0°C/Cu10 WEED: -20.0 to 250.0°C/
Cu10 BAILEY: -20.0 to 250.0°C
9 Measured using a measurement current of approximately 1 mA at 1-V range. Threshold level is approximately 0.8 V.
DI
RTD
(Measurement
current: 1 mA)
Thermocouple
(excludes RJC accuracy, when burnout
is OFF)
Measurement Accuracy
Integration Time: 1.67 ms
±(0.1% of rdg + 25digits)
±(0.1% of rdg + 10digits)
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
RTD
(Measurement
current: 2 mA)
Measurement
Range
Type