4-9
IM MX100-01E
Specification
The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).
60 mV (high res.)
1 V
6 V (high res.)
PLATINEL
PR40-20
NiNiMo
WRe3-25
W/WRe26
Type-N(AWG14)
Pt100 (high
noise resistance)
JPt100 (high
noise resistance)
Cu10 at 20°C
alpha = 0.00392
Cu10 at 20°C
alpha = 0.00393
Cu25 at 0°C
alpha = 0.00425
Cu53 at 0°C
alpha = 0.00426035
Cu100 at 0°C
alpha = 0.00425
Pt25(JPt100*1/4)
Cu10 GE
(high resolution)
Cu10 L&N
(high resolution)
Cu10 WEED
(high resolution)
Cu10 BAILEY
(high resolution)
Pt100 (noise
resistance)
JPt100 (noise
resistance)
0.000 to 60.000 mV
-1.0000 to 1.0000 V
0.0000 to 6.0000 V
0.0 to 1400.0°C
0.0 to 1900.0°C
0.0 to 1310.0°C
0.0 to 2400.0°C
0.0 to 2400.0°C
0.0 to 1300.0°C
-200.0 to 600.0°C
-200.0 to 550.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-50.0 to 150.0°C
-50.0 to 150.0°C
-200.0 to 550.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 250.0°C
-200.0 to 250.0°C
±(0.05% of rdg + 20digits)
±(0.05% of rdg + 2digits)
±(0.05% of rdg + 20digits)
±(0.05% of rdg + 1°C)
±(0.05% of rdg + 2.5°C)
Except 300 to 700°C: ±6°C
less than 300°C: not guaranteed
±(0.05% of rdg + 0.7°C)
±(0.05% of rdg + 2°C)
Except 0 to 200°C: ±2.5°C
2000°C or higher:
±(0.05% of rdg + 4°C)
±(0.05% of rdg + 2°C)
Except 100 to 300°C: ±4°C
less than 100°C: not guaranteed
±(0.05% of rdg + 0.7°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 0.7°C)
±(0.1% of rdg + 0.5°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 0.5°C)
±(0.1% of rdg + 0.7°C)
±(0.05% of rdg + 0.3°C)
1 µV
100 µV
100 µV
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
0.1°C
±(0.1% of rdg + 100 digits)
±(0.1% of rdg + 10 digits)
±(0.1% of rdg + 100 digits)
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Not measurable
Input
Measurement Accuracy
Integration Time: 16.67 ms
or More
Highest
Resolution
(1 Digit)
Rated
Measurement
Range
Measurement Accuracy
Integration Time: 1.67 ms
Measurement
Range
Type
DC voltage
RTD
(Measurement
current: 1 mA)
Thermocouple
(excludes RJC accuracy,
when burnout is OFF)
RTD
(Measurement
current: 2 mA)
Measurementinterval,integrationtime,andlter:
Measurement
Interval
10 ms
1.67 ms
16.67 ms
20 ms
Auto
Integration
Time
Filter
Rejected Noise and Notes
50 ms
Rectangular
100 ms
200 ms
500 ms
1 s
2, 5,10, 20, 30, 60 s
36.67 ms
100 ms
200 ms
Trapezoidal
Rectangular
Cos
600 Hz and its integer multiples, temperature measurement not possible
60 Hz and its integer multiples
50 Hz and its integer multiples
Automatically detects the power supply frequency and set 16.67 or 20 ms
50 Hz or 60 Hz and their integer multiples
10 Hz and its integer multiples
Fc = 5-Hz low-pass filter
Reference junction compensation:
Switch between external and internal on each channel, remote RJC
function available (see page 4-3)
Reference junction compensation accuracy:
When measuring temperature greater than or equal to 0°C and when
the temperature of the input terminal is balanced
Type R, S, W:
±1°C
Type K, J, E, T, N, L, U:
±0.5°C
Type N(AWG14), PLATINEL, NiNiMo, WRe3-25, W/WRe26:
±1°C
Note: Theinternalreferencejunctioncompensationisxedto0°C
for type B and PR40-20