4-16
IM MX100-01E
4.6 6-CH, Medium-Speed 4-Wire RTD Resistance
Input Module Specifications
Style number: S2
Type of measurement: DC voltage, 4-wire RTD, 4-wire resistance, DI (contact, LEVEL)
Number of inputs: 6
Input method: Floating unbalanced input, insulation between channels
A/Dresolution: 16bits(±20000/±6000)
Measurement range and accuracy:
The accuracy applies to standard operating conditions:
Ambienttemp:23±2°C,ambienthumidity:55±10%RH,supply
voltage:90to250VAC,powerfrequency:50/60Hz±1%,warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
20 mV
60 mV
200 mV
2 V
6 V
20 V
100 V
Pt100
1
JPt100
1
Pt100 (high res.)
JPt100 (high res.)
Ni100 SAMA
2
Ni100 DIN
2
Ni120
3
Pt50
Cu10 GE
4
Cu10 L&N
4
Cu10 WEED
4
Cu10 BAILEY
4
J263B
Pt500
6
Pt1000
6
20 Ω
(Measurement
current 1 mA)
200 Ω
(Measurement
current 1 mA)
2 kΩ
(Measurement
current 0.25 mA)
Level
Contact input
DC voltage
Resistance
7
–20.000 to 20.000 mV
–60.00 to 60.00 mV
–200.00 to 200.00 mV
–2.0000 to 2.0000 V
–6.000 to 6.000 V
–20.000 to 20.000 V
–100.00 to 100.00 V
–200.0 to 600.0°C
–200.0 to 550.0°C
–140.00 to 150.00°C
–140.00 to 150.00°C
–200.0 to 250.0°C
–60.0 to 180.0°C
–70.0 to 200.0°C
–200.0 to 550.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
0.0 to 300.0K
–200.0 to 600.0°C
–200.0 to 600.0°C
0.000 to 20.000 Ω
0.00 to 200.00 Ω
0.0 to 2000.0 Ω
Vth = 2.4 V
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 2°C)
±(0.05% of rdg + 0.3K)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 7digits)
±(0.05% of rdg + 3digits)
±(0.05% of rdg + 3digits)
Threshold level accuracy: 0.1 V
ON for 100 Ω or less, OFF for 1 kΩ or more
5
±(0.1% of rdg + 25digits)
±(0.1% of rdg + 10digits)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5K)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 25digits)
±(0.1% of rdg + 15digits)
±(0.1% of rdg + 10digits)
Input
Measurement Accuracy
Integration Time: 16.67 ms
or More
Rated
Measurement
Range
Measurement Accuracy
Integration Time: 1.67 ms
Measurement
Range
Type
1 Pt50: JIS C 1604-1981, JIS C 1606-1986/Pt100: JIS C 1604-1989, JIS C 1606-1989, IEC 751, DIN IEC 751/
JPt100: JIS C 1604-1981, JIS C 1606-1989
2 SAMA/DIN
3 McGRAW EDISON COMPANY
4 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/
Cu10 BAILEY: –20.0 to 250.0°C
5 Measured using a measurement current of approximately 10 µA at 200 V range. Threshold level is approximately 0.1 V.
6. The Pt500 resistance table is Pt100 × 5, and the resistance table for Pt1000 is Pt100 × 10.
7. 4-wire RTD, 4-wire resistance
Highest
Resolution
(1 Digit)
1 µV
10 µV
10 µV
100 µV
1 mV
1 mV
10 mV
0.1°C
0.01°C
0.1°C
0.1°C
0.1°C
0.1K
0.1°C
0.001Ω
0.01Ω
0.1Ω
DI
RTD
7
(Measurement
current:
0.25 mA)
RTD
7
(Measurement
current: 1 mA)