EasyManua.ls Logo

YOKOGAWA MX100 - 4.6 6-CH, Medium-Speed 4-Wire RTD Resistance Input Module Specifications

YOKOGAWA MX100
126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4-16
IM MX100-01E
4.6 6-CH, Medium-Speed 4-Wire RTD Resistance
Input Module Specifications
Style number: S2
Type of measurement: DC voltage, 4-wire RTD, 4-wire resistance, DI (contact, LEVEL)
Number of inputs: 6
Input method: Floating unbalanced input, insulation between channels
A/Dresolution: 16bits(±20000/±6000)
Measurement range and accuracy:
The accuracy applies to standard operating conditions:
Ambienttemp:23±2°C,ambienthumidity:55±10%RH,supply
voltage:90to250VAC,powerfrequency:50/60Hz±1%,warm-
up time: at least 30 minutes, without adverse conditions such as
vibrations.
20 mV
60 mV
200 mV
2 V
6 V
20 V
100 V
Pt100
1
JPt100
1
Pt100 (high res.)
JPt100 (high res.)
Ni100 SAMA
2
Ni100 DIN
2
Ni120
3
Pt50
Cu10 GE
4
Cu10 L&N
4
Cu10 WEED
4
Cu10 BAILEY
4
J263B
Pt500
6
Pt1000
6
20 Ω
(Measurement
current 1 mA)
200 Ω
(Measurement
current 1 mA)
2 kΩ
(Measurement
current 0.25 mA)
Level
Contact input
DC voltage
Resistance
7
–20.000 to 20.000 mV
–60.00 to 60.00 mV
–200.00 to 200.00 mV
–2.0000 to 2.0000 V
–6.000 to 6.000 V
–20.000 to 20.000 V
–100.00 to 100.00 V
–200.0 to 600.0°C
–200.0 to 550.0°C
–140.00 to 150.00°C
–140.00 to 150.00°C
–200.0 to 250.0°C
–60.0 to 180.0°C
–70.0 to 200.0°C
–200.0 to 550.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
–200.0 to 300.0°C
0.0 to 300.0K
–200.0 to 600.0°C
–200.0 to 600.0°C
0.000 to 20.000 Ω
0.00 to 200.00 Ω
0.0 to 2000.0 Ω
Vth = 2.4 V
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 5 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 2°C)
±(0.05% of rdg + 0.3K)
±(0.05% of rdg + 0.3°C)
±(0.05% of rdg + 7digits)
±(0.05% of rdg + 3digits)
±(0.05% of rdg + 3digits)
Threshold level accuracy: 0.1 V
ON for 100 Ω or less, OFF for 1 kΩ or more
5
±(0.1% of rdg + 25digits)
±(0.1% of rdg + 10digits)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 1.5K)
±(0.1% of rdg + 1.5°C)
±(0.1% of rdg + 25digits)
±(0.1% of rdg + 15digits)
±(0.1% of rdg + 10digits)
Input
Measurement Accuracy
Integration Time: 16.67 ms
or More
Rated
Measurement
Range
Measurement Accuracy
Integration Time: 1.67 ms
Measurement
Range
Type
1 Pt50: JIS C 1604-1981, JIS C 1606-1986/Pt100: JIS C 1604-1989, JIS C 1606-1989, IEC 751, DIN IEC 751/
JPt100: JIS C 1604-1981, JIS C 1606-1989
2 SAMA/DIN
3 McGRAW EDISON COMPANY
4 Guaranteed accuracy range Cu10 GE: –84.4 to 170.0°C/Cu10 L&N: –75.0 to 150.0°C/Cu10 WEED: –20.0 to 250.0°C/
Cu10 BAILEY: –20.0 to 250.0°C
5 Measured using a measurement current of approximately 10 µA at 200 V range. Threshold level is approximately 0.1 V.
6. The Pt500 resistance table is Pt100 × 5, and the resistance table for Pt1000 is Pt100 × 10.
7. 4-wire RTD, 4-wire resistance
Highest
Resolution
(1 Digit)
1 µV
10 µV
10 µV
100 µV
1 mV
1 mV
10 mV
0.1°C
0.01°C
0.1°C
0.1°C
0.1°C
0.1K
0.1°C
0.001Ω
0.01Ω
0.1Ω
DI
RTD
7
(Measurement
current:
0.25 mA)
RTD
7
(Measurement
current: 1 mA)

Table of Contents

Related product manuals