4-17
IM MX100-01E
Specification
The inputs indicated below can be used on MXLOGGER or the MX100/DARWIN API
(software sold separately).
60mV (high res.)
1V
6V (high res.)
Cu10 at 20°C
alpha=0.00392
Cu10 at 20°C
alpha=0.00393
Cu25 at 0°C
alpha=0.00425
Cu53 at 0°C
alpha=0.00426035
Cu100 at 0°C
alpha=0.00425
Pt25(JPt100/4)
*
Cu10 GE
(high resolution)
Cu10 L&N
(high resolution)
Cu10 WEED
(high resolution)
Cu10 BAILEY
(high resolution)
0.000 to 60.000 mV
-1.0000 to 1.0000 V
0.0000 to 6.0000 V
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-50.0 to 150.0°C
-50.0 to 150.0°C
-200.0 to 550.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
-200.0 to 300.0°C
±(0.05% of rdg + 20 digits)
±(0.05% of rdg + 2 digits)
±(0.05% of rdg + 20 digits)
±(0.1% of rdg + 2°C)
±(0.1% of rdg + 0.5°C)
±(0.05% of rdg + 0.3°C)
±(0.1% of rdg + 0.5°C)
±(0.1% of rdg + 2°C)
1µV
100µV
100µV
1°C
0.1°C
0.1°C
0.1°C
0.1°C
±(0.1% of rdg + 100 digits)
±(0.1% of rdg + 10 digits)
±(0.1% of rdg + 100 digits)
±(0.2% of rdg + 5°C)
±(0.2% of rdg + 2°C)
±(0.1% of rdg + 1.5°C)
±(0.2% of rdg + 2°C)
±(0.2% of rdg + 5°C)
DC voltage
Input
Measurement Accuracy
Integration Time: 16.67 ms
or More
Rated
Measurement
Range
Measurement Accuracy
Integration Time: 1.67 ms
Measurement
Range
Type
Highest
Resolution
(1 Digit)
4-wire RTD
(Measurement
current: 1 mA)
Measurementinterval,integrationtime,andlter:
Measurement
Interval
1.67 ms
16.67 ms
20 ms
Auto
36.67 ms
100 ms
200 ms
Integration
Time
Filter
Rejected Noise and Notes
Rectangular
100 ms
200 ms
500 ms
1 s
2 s
5 s
10, 20, 30, 60 s
Trapezoidal
Rectangular
Cos
600 Hz and its integer multiples, temperature measurement not possible
60 Hz and its integer multiples
50 Hz and its integer multiples
Automatically detects the power supply frequency and set 16.67 or 20 ms
50 Hz, 60 Hz and their integer multiples
10 Hz and its integer multiples
Fc = 5-Hz low-pass filter
* When the measurement interval is 100 ms and 200 ms, measured values may fluctuate since power supply frequency
noise is not rejected. In such cases, set the measurement interval to 500 ms or more.
Maximum input voltage:
DCvoltageat1-Vrangeorless,RTD,resistance,andDI(contact):±10
VDC (continuous)
Other measurement ranges:
±120 VDC (continuous)
Normal-mode voltage:
DC voltage, DI (LEVEL): 1.2 times the range rating or less (50/60 Hz, peak
value including the signal component). The RTD and resistance ranges
indicate the voltage conversion value when measurement current applied.)
Resistance2kΩ,RTD100Ω,500Ω,1000Ωsystems:50mVpeak
Resistance200Ω,RTD10Ω,25Ω,50Ωsystems:10mVpeak
Resistance20Ω:4mVpeak
Normal-mode rejection ratio:
RTD and registance ranges are voltage conversion values when the
measuredcurrentisowing.
Forintegrationtimeof16.67msormore:40dBormore(50/60Hz±0.1%)
For integration time of 1.67 ms: 50/60 Hz is not rejected
Common-mode voltage: 600 VACrms (50/60 Hz), reinforced (double) insulation
Common-mode rejection ratio:
(50/60Hz±0.1%,500Ωunbalanced,betweentheminusterminaland
ground, RTD, and resistance ranges are voltage conversion values when
themeasuredcurrentisowing.)
When integration time is 16.67 ms or more: 120 dB or more
When integration time is 1.67 ms or more: 80 dB or more
Common-mode voltage between channels:
AC voltage, DI: 120 VACrms (50/60 Hz)
RTD, resistance: 50 VACrms (50/60 Hz)
Noiserejection: RejectionbytheintegratingA/Dconverterandtheuseoflowpasslters
4.6 6-CH, Medium-Speed 4-Wire RTD Resistance Input Module Specifications