4-18
IM MX100-01E
Inputresistance DCvoltageat1Vrangeorless:10MΩormore
DCvoltageat2Vrangeormore:Approx.1MΩ
Insulationresistance: 20MΩormore(500VDC)betweentheinputandearth
Input bias current: 10 nA or less
Withstand voltage: 1000 VAC rms (50/60 Hz) for one minute between input terminals (DC
voltage and DI)
620 VAC rms (50/60 Hz) for one minute between input terminals (RTD
and resistance)
3700 VAC rms (50/60 Hz) for one minute between the input terminal
and earth terminal
Input signal source resistance:
DCvoltage:2kΩorless
Resistance,RTDranges:10Ωorlessperwire(commontoallranges)
Shunt capacity: 0.01 mF or less (when using the RTD or resistance range)
Power consumption:Approximately 1.2 W
External dimensions:Approximately 57 (W) × 131 (H) × 151 (D) mm (including the terminal cover)
Weight: Approximately 0.5 kg
Terminal type: Clamp, terminal board is detachable
Applicable wire size:0.14 to 1.5 mm
2
(AWG26 to 16)
Effects of Operating Conditions
The specifications below apply when the integration time is 16.67 ms or more.
Warm-up time: At least 30 minutes after power-up
Effects of ambient temperature:
The effect received by changes in 10°C increments of the ambient
temperatureis: ±(0.05%ofrdg+0.05%ofrange)orless
However,±(0.2%ofrange+1digit)forCu10Ω
Effectsofpoweructuation: MeetstheaccuracyspecicationsforACpowersupplyin
the range 90 to 132 V and 180 to 250 V
Effectsofmagneticeld: TheuctuationinexternalmagneticeldsforAC(50/60
Hz)400A/mis±(0.1%ofrdg+10digits)orless
Effects from the signal source resistance:
EffectsonDCvoltagefromauctuationof1kΩinthesignalsourceresistance
1Vrangeorless:±10mVorless
2Vrangeorhigher:±0.15%ofrdgorless
RTD: Fluctuationsresultingfrom10Ωchangesperwire:
1000Ωand100Ωsystems:±0.1°Corless
Otherthan1000Ωand100Ωsystems:±1.0
°C or less
Resistance: Fluctuationsresultingfrom10changesperwire:±1digitorless
Effect of position: Horizontal with the feet at the bottom is the rule.
Effectsofvibration: Theuctuationthatresultsbyapplyingasinusoidalvibration
along all three axis at a frequency between 10 to 60 Hz and an
acceleration of 0.2 m/s
2
is±(0.1%ofrdg+1digit)orless
External Dimensions
Unit: mm
57
150.8
131
If not specified, the tolerance is ±3%.However,incasesoflessthan10mm,thetoleranceis±0.3mm.
4.6 6-CH, Medium-Speed 4-Wire RTD Resistance Input Module Specifications

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the YOKOGAWA MX100 and is the answer not in the manual?

YOKOGAWA MX100 Specifications

General IconGeneral
BrandYOKOGAWA
ModelMX100
CategoryData Loggers
LanguageEnglish

Related product manuals