Operation Manual
Precautions for Use
* Warning
CARELESS OR IMPROPER USE OF X-RAY EQUIPMENT CAN BE EXTREMELY
HAZARDOUS.
1. This product is only to be used in the radiaton control area.
2. During image acquisition or positioning, the rotation arm unit will rotate around the
patient. To avoid the accident, look out for patient's safety at all time. In case of any
dangerous situation occurs, immediately stop the operation.
3. During image acquisition or positioning, the rotating arm unit will rotate around the
patient. To avoid the accident, ask the patient to refrain from moving until rotating arm
unit stops the movement.
4. Ensure that the power switch is OFF (O) if you do not intend to operate the equipment
for a long period.
5. Modicationofthedeviceisprohibited.Donotremoveanycoversorcableson
system.
6. Electromagnetic wave
This medical device could malfunction due to electromagnetic waves. Instruct persons
in your fascility to turn off the power of electronic devices such as mobile phones,
transceivers, radio control transmitters, etc.
7. Be sure to plug a medical ground cable into a ground outlet prior to use of X-era
Smart. Do not plug the cable into any other kind of outlet than that for exclusive use.
8. Donotconnectanydevicesotherthanthespeciedforthisequipment.Also,donot
installsoftwareotherthanthespeciedtothePCconnectedtothesystem.
9. Donotinstallthelmprocessorinthesameroomastheequipment.Thegas
generatedbythelmprocessormaycorrodetheequipmentandcauseserious
damage.
10. Do not look directly into the positioning beam. The power output (Class 2) is at the
level which adverse effect on the eyes can be avoided by blinking. However caution
should be taken for maximum safety.
11. Do not apply any heavy load or shock on the equipment, e.g. suspending from the chin
rest or sliding mechanism, otherwise the equipment may be damaged.
12. Be sure to keep the infants and small children away from the equipment unless
required for radiographic purpose.
13. Do not give high impact or excessive external load to this product. It may cause
malfunction.