1.7 Cooling Time
The equipment has incorporated a variable-cooling time function. X-ray radiographic may
be repeated consecutively three or four times after it has not been used for a long time;
however, the X-ray tube head, once reaching a given temperature, should be cooled for a
given time depending on the radiographic conditions as shown in the following table.
The cooling time of the equipment are as described in the following table.
Panoramic mode
TubeVoltage[kV] TubeCurrent[mA]
Exposure Time
[sec]
Cool.Time[sec]
Max. exposure/hr
[No.oftimes]
Con.1 73 6.3 16 352 10
Con.2 73 6.3 14 308 12
Con.3 79 6.3 16 383 10
Con.4 79 6.3 14 335 11
TMJ mode
TubeVoltage[kV] TubeCurrent[mA]
Exposure Time
[sec]
Cool.Time[sec]
Max. exposure/hr
[No.oftimes]
Con.1 73 10 8 232 15
Con.2 79 10 8 256 13
Cephalometric and Carpus radiography (*)
TubeVoltage[kV] TubeCurrent[mA]
Exposure Time
[sec]
Cool.Time[sec]
Max. exposure/hr
[No.oftimes]
Con.1 82 10 8 320 12
Con.2 82 10 10 400 9
3D X-ray image acquisition cycle (**)
TubeVoltage[kV] TubeCurrent[mA]
Exposure Time
[sec]
Cool.Time[sec]
Max. exposure/hr
[No.oftimes]
Con.1 82 4.0 11.5 178 19
Con.2 82 4.0 23 355 9
* For equipment with Cephalometric
** For equipment with 3D function