Maintenance Services
129
Operation Manual
No. Check that: Timing Remarks
11
X-ray irradiator (Head unit) and the image reception
area (Sensor unit) are clean with no dent, dirt or
dust.
pre-operation
The user is responsible for performing the regular checks listed above.
For more information on how to perform regular checks, see "7.4 Performing Regular Check
Services".
7.3 Periodical Check Services
The periodical checks listed below are required on the equipment.
Periodical check list
No. Check that: Timing Remarks
1
Supplypower(inputpower:90-132Vor198-264V)
is supplied.
every 6 months
2
The protective earth line is correctly connected to
ground.
every 6 months
3 QA (Quality Assurance) every 6 months
The user is responsible for performing the periodical checks listed above.
The periodical checks can be outsourced to a subcontractor. In such case, the checking
items are differ from the above. Please see 7.3.1.
For more information on how to perform above periodical checks, see "7.4 Performing
Regular Check Services".
7.3.1 Outsourcing the Periodical check services
The periodical check services can be outsourced to a subcontractor. The main check
services are listed below. The details are described in the “maintenance and inspection list”
kept by your dealer. Please ask the dealer for further information.
List of Periodical check by subcontractor
No. Check that: Timing Remarks
1
Supplypower(inputpower:90-132Vor198-264V)
is supplied.
every 6 months
2
The protective earth line has been connected to
ground.
every 6 months
3 X-ray output is normal. every 6 months
4 Operational safety, e.g., in arm rotating, is assured. every 6 months
5 All the sensors operate properly. every 12 months
6 Electrical safety is assured. every 12 months
7.3.2 Archiving records on Periodical check
To ensure the reliability and safety of X-era Smart, be sure to archive the result records on
regular check and scheduled check services.