Trouble Shooting
126
Operation Manual
Error code Description Action
Er75 (*) Cephalometric relay board communication error
Press the Reset switch on the
control panel to recover from the
error. If this error persists, contact
your dealer.
Er79 (**) 3D sensor slide position error
Er80 (*) Collimator switching limit sensor error
Er81 (*)(**) Collimator switching position error
Er82
Sensor (detachable) is not attached to Docking Station
(Panorama) when acquiring panoramic image.
The sensor unit (detachable) is not detached to the docking
station unit (Panoramic) when acquiring a 3D image.(**)
Er83 (*)
Sensor (detachable) is not attached to Docking Station
(Cephalo) when acquiring cephalometric image.
Er84 (*)
Motor movement of the Cephalometric unit did not
complete properly.
* For equipment with Cephalometric
** For equipment with 3D function
Warning message code list
Warning Code Description Action
FL01
Notice of service maintenance services.
Total No. of Exposure has exceeded 2,000.
It is recommended that you will perform regular
checks. Contact your dealer.
FL02
The exposure switch has been pressed
during selecting Exposure conditions.
It will be cancelled by the reset switch on
the Control panel. Contact your dealer if it
happens frequently.
FL03
The exposure switch has been pressed
during X-ray cooling time period.
FL05(**)
Theheadxatorfor3Dimageacquisitionis
attached other than in 3D image acquisition
mode.
Pressing the reset switch on the control panel
releasestheerror.Removetheheadxatorfor
3D image acquisition.
FL06(**)
Theheadxatorfor3Dimageacquisitionis
not attached in 3D image acquisition mode.
Pressing the reset switch on the control panel
releasestheerror.Attachtheheadxatorfor
3D image acquisition.
** For equipment with 3D function
6.2 Contact Procedures at the Time of Failure
If any problem persists even though the appropriate corrective measure described above has
been taken, contact your dealer and provide information on:
1. Product name (X-era Smart) and its serial number (**-****).
2. Operation, which might cause the failure (as detailed as possible).
3. Current condition
e.g.:Anerrormessage○○isdisplayed.Thearmunitorslidingmechanismhasstopped
with .......