2.5.6 X-ray Exposure
* Caution
• Onlytrainedandqualiedpersonshouldtakeradiographicphotos.X-rayexposure(or
imageacquisition)byunqualiedpersonisprohibitedbylaw.
• To minimize the X-ray exposure, properly position the patient.
• Instruct the patient not to move during the image acquisition. When the patient moves
during the image acquisition, the image may get blurred and the quality low. Especially
for 3D oral image acquisition, advise the patient to stay still until the whole process of
the acquisition is completed.
• Before 3D image acquisition, make sure that the panoramic sensor is detached from the
sensor part (detachable). If image acquisition is proceeded with the panoramic sensor
attached, 3D image cannot be acquired properly, and the patient may get accidentally
injured due to the interference with the equipment.
• Before 3D image acquisition, make sure that the head part and movable sensor part are
not interfering with the patient's head.
Starting X-ray Exposure
After the operations described in "From the control panel:" P.56 or , "From the Image
Creator :" P.59 make sure that the Image Creator's Image Construction Screen is under
Acquisition standby status.