EasyManua.ls Logo

YOSHIDA XP73 - Main Specification; Specification

Default Icon
196 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MainSpecication
135
Operation Manual
11 MainSpecication
11.1 Specication
Item Specications
Equipment type
Digital panoramic, tomographic X-ray diagnostic equipment, arm-type
X-ray CT diagnostic equipment for dental use.
Product name (Model name) X-era Smart
Type number XP73
Operating mode Continuous operation with intermittent loading
High-voltage generator DCrectication
X-ray tube focal spot size 0.5 mm × 0.5 mm
X-ray tube cooling method Oil cooling
Nominal maximum electric power
(combination of X-ray tube voltage
and tube current at maximum output)
0.82kW(82kV,10mA)
Tube voltage 58,60,63,65,68,71,73,76,79,82kV(±5%)
Tube current 2.0, 2.5, 3.2, 4.0, 5.0, 6.3, 8.0, 10 mA (±15%)
Exposure time
Panoramic
Panorama 8, 14, 16 sec
±(10% + 1ms)
Child 6.4, 11.2, 12.8 sec
TMJ 4 sec × 2
Single
0.10, 0.12, 0.16, 0.20, 0.25,
0.32, 0.40, 0.50, 0.63, 0.80,
1.00, 1.20, 1.60, 2.00, 2.50,
3.20 sec
Cephalometric / Carpus
image acquisition (*)
8, 10 sec
3D Oral image radiography
(**)
11.5 seconds x 2
3D Dental image acquisition
(**)
11.5 seconds
Photographing direction
Panorama, TMJ and 3D Oral: clockwise/counterclockwise, 3D Dental:
clockwise rotation
Imagemagnicationfactor
Panoramic image acquisition,
TMJ image acquisition
1.2 - 1.29
Cephalometric / Carpus
image acquisition (*)
1.1
Filter type
Add Filter 1.5 mmAl
Exposureoriceplate0.17mmAl
Criteria:TubeVoltage/Current=85kV/12mA
Totalltration 2.5mmAlequivalentorover(atTubevoltage:82kV)
Leakage dose 0.88 mGy/h or less
Leakage dose calculation standards Tubevoltage82kV,tubecurrent10mA,Dutycycle1:40
Duty cycle 1:40