* Caution
14. To minimize the X-ray dose irradiated on the patient, be sure to position the patient
correctly prior to image acquisition.
15. To prevent the patient from infection, always use a new bite block for each patient and
wipe it with alcohol-dempened cotton prior to use.
16. Correctly connect the PC protective earth terminal to ground.
17. Do not rotate the arm unit manually, as the patient body may be caught by the arm unit
or a failure may occur in the equipment.
18. Once the arm unit is positioned, be sure to return the positioning mirror to its original
position.
19. Becarefulnottogetyourngerscaughtintheequipmentduringtheoperation,such
as rotating, opening and closing, moving back and forth, of the ear rod and nose
support of the Cephalometric unit.
20. Do not touch the Sensor unit and the Second Slit unit while the Cephalometric Sensor
isoperating.Itisdangerousasyourngersorhandsmaygetcaughtandcause
injured.
21. Do not apply force to the Sensor unit. It may cause failure in the equipment.
22. Be sure to open or close the ear rod of the Cephalometric unit by its knob without
applying excessive force. Otherwise the patient may be injured.
23. Be sure to connect the additional protective earth conductor of the equipment.
24. Beware not to touch the equipment while the movable 3D sensor is in motion. The
ngerorhandmightgetcaughtandinjured.
25. Look out for the safety of the patient when moving the nose support of the
Cephalometric unit up, down, back or forth.
26. Advise the patient to stay still while an image is being acquired. If the patient moves,
it causes image blurring and degrades the image quality.Especially for 3D oral image
acquisition, advise the patient to stay still until the whole process of the acquisition is
completed.
27. Setupthemonitorwhereisnotaffectedbythereectionoflightusedinsideoroutside
the room and ensure you get the most suitable viewing condition and quality of the
image.
28. Do not operate this equipment in case of earthquake occurrence. After an earthquake,
check to see if the equipment operates properly to ensure the safety of the patient.
29. DonotplaceanyobjectsinsidetheX-rayeldoftheequipment.
30. Whenattachinganddetachingthesensor,rmlyholditbytheknobandmakesure
you do not drop it. Dropping the sensor may cause failure of the equipment.
31. The sensor unit is equipped with a shock sensor. The shock sensor records any
excessive impact incurred, such as when the equipment falls down, and the equipment
will not qualify as being warrantable. Exercise caution in handling the equipment.