User Program Mode
121
Operation Manual
Descriptions of User Program Mode
Program No. Description Setting Default
Prog No.02
Specieswhetheran
audible alarm sounds
when any button is
pressed on the touch
panel.
Audible alarm
On
On: an alarm sounds Off: no alarm sounds
Prog No.03
Specieswhether
the arm automatically
moves to the patient
positioning point after
image acquisition.
Patient positioning point (after acquisition)
On
On:specied Off:notspecied
Prog No.04
Speciespower-on
behavior
Power-on behavior
Entr
Off: no auto
behavior
Entr: patient
positioning
point
Rest: auto reset
point
Prog No.05
Speciesthe
correction level of
cervical spine for
image acquisition
Cervical spine correction
Lo
Off: no correction Hi: step 2 Lo: step 1
Prog No.06
Specieswhethera
service maintenance
message appears
once 2,000 th
exposure has
nished.
Service maintenance message
No
Yes: no message appears No: message appears
Prog No.07
Resets the settings
in memory to their
defaults.
User program settings
No
Yes: resets to defaults No: existing settings retained
Prog No.09
Stores the initial state
in memory at power-
on.
Initialization at power-on
No
Yes: Initialization at power-on No: not stored in memory
Prog No.11
Specieswhetheran
audible alarm sounds
when an error occurs.
Audible alarm sound
On
On: an alarm sounds Off: no alarm sounds
Prog No.22*
Sets the irradiation
conditions of the
AUTO mode in 3D
image acquisition.
Irradiation conditions in 3D exposure.
2
1:82kV,2.0mA 2:82kV,4.0mA 3:82kV,6.3mA
* For equipment with 3D function
Suppl. : During setting the mode in the User Program Mode, any operation from the
exposure switch or control panel is inactivated.