Directions for Use
73
Operation Manual
Adjusting the forehead support
Slide the forehead support in contact
with the patient’s head by gripping its
base and then close it. With the Auto
setting, the tube voltage appropriate for
the patient is automatically set.
At the end of adjustment, turn on the
positioningbeamagaintoconrmthat
the beam is right on the proper point.
* Caution
• Do not open or close the head support forcedly without gripping its base. Otherwise, the
patient may be injured.
Turning off positioning beam
Press the Beam switch to turn off the positioning beam. Return the positioning mirror to its
original position.
* Caution
• Oncepositioninghasbeensuccessfullynished,besuretoreturnthepositioningmirror
to its original position.
* Caution
• After the positioning is completed, be sure to return the positioning mirror to its original
position.