Page #187 background imageLoading...
Page #187 background image
187
Checking the meter 7
Did you perform the control test according to
the User’s Manual?
yes
Did you wait for the application prompt to be
displayed before applying control solution?
yes
Did you apply the control solution immedi-
ately after opening the applicator?
yes
Did you enter the correct color? yes
Did you perform the test within the correct
temperature range (10–40 °C)?
yes
Has the expiry date for the control solution
been reached (see label on the bottom of the
vial next to the symbol
)?
no
Did you read the section “Storage and stabil-
ity of control solutions” in the package insert
for the control solution?
yes
Did you observe the proper storage condi-
tions for the meter, test cassette and control
solutions (see Chapter 15 and the test cas-
sette and control solution package inserts)?
yes
If your meter has been dropped, that can also lead to
control test results which are not within the concentra-
tion range displayed and for which Control test not OK
is displayed. In this case, you should also call
Accu-Chek Customer Care.

Table of Contents

Other manuals for Accu-Chek Mobile

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Mobile and is the answer not in the manual?

Accu-Chek Mobile Specifications

General IconGeneral
Measurement TechnologyPhotometric
Sample TypeCapillary whole blood
Sample Size0.3 µL
Test Time5 seconds
Battery Type2 AAA batteries
ConnectivityUSB
Dimensions121 x 63 x 20 mm
Weight129 g (with batteries)
DisplayLCD
Memory Capacity2000 results
Measuring Range0.6 - 33.3 mmol/L
Operating Temperature10 - 40°C
Operating Humidity10 to 90% RH
Hematocrit Range20 to 70%
Test Strip TechnologyNo test strips required (uses a test cassette)