Checking the meter7
Control test not OK – possible sources of error
If Control test not OK ... Repeat control test is dis-
played, the result obtained is not within the concentra-
tion range displayed. Check the points listed on the
next page. If your answer is different from the answer
given, make the respective corrections and repeat the
test. If you have taken all of these points into account
and Control test not OK is still displayed, contact
Accu-Chek Customer Care.
If Control test not OK Repeat control test is
displayed, it is no longer certain that the meter
and test cassette are functioning properly. Blood
glucose tests may then deliver incorrect results.
Incorrect results can cause the wrong therapy
recommendation to be made and thus produce
serious adverse health effects.

Table of Contents

Other manuals for Accu-Chek Mobile

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Mobile and is the answer not in the manual?

Accu-Chek Mobile Specifications

General IconGeneral
Measurement TechnologyPhotometric
Sample TypeCapillary whole blood
Sample Size0.3 µL
Test Time5 seconds
Battery Type2 AAA batteries
ConnectivityUSB
Dimensions121 x 63 x 20 mm
Weight129 g (with batteries)
DisplayLCD
Memory Capacity2000 results
Measuring Range0.6 - 33.3 mmol/L
Operating Temperature10 - 40°C
Operating Humidity10 to 90% RH
Hematocrit Range20 to 70%
Test Strip TechnologyNo test strips required (uses a test cassette)