217
Testing and storage conditions 15
Testing and storage conditions15
Make sure that the conditions listed on the following
pages are met, so that the meter and lancing device
work reliably and you obtain accurate results.
15.1 Temperature and humidity
Without
batteries and without a test cassette, the
meter should be stored at between -25 and +70 °C.
With
batteries and without a test cassette, the
meter should be stored at between -10 and +50 °C.
At temperatures above +50 °C, the batteries
could leak and damage the meter.
At temperatures below -10 °C, the batteries
do not have enough power to keep the
internal clock functioning.
With
batteries and with a test cassette, the meter
should be stored at between +2 and +30 °C.
Store the meter in a place with a relative humidity
between 15 and 93 %.
The temperature and humidity conditions under
which the meter can operate are +10 to 32 °C at
15 to 85 % relative humidity, including the following
exteme conditions: +32 °C at 85 % relative humidity
and +40 °C at 45 % relative humidity.
Sudden changes in temperature can cause
condensation to form in or on the meter. If this
happens, do not turn the meter on. Make sure
that it returns to ambient temperature slowly.
Never store the meter in a room that is likely to
harbour condensation (e.g. bathroom).
If the temperature is between +8 and +10 °C or
between +40 and +42 °C, the meter will still allow
you to perform a test. However, the following warn-
ing will appear (see also Chapter 16.2).

Table of Contents

Other manuals for Accu-Chek Mobile

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Mobile and is the answer not in the manual?

Accu-Chek Mobile Specifications

General IconGeneral
Measurement TechnologyPhotometric
Sample TypeCapillary whole blood
Sample Size0.3 µL
Test Time5 seconds
Battery Type2 AAA batteries
ConnectivityUSB
Dimensions121 x 63 x 20 mm
Weight129 g (with batteries)
DisplayLCD
Memory Capacity2000 results
Measuring Range0.6 - 33.3 mmol/L
Operating Temperature10 - 40°C
Operating Humidity10 to 90% RH
Hematocrit Range20 to 70%
Test Strip TechnologyNo test strips required (uses a test cassette)

Related product manuals