Cleaning the meter and lancing device 11
The inside
You only need to clean the inside of the meter if:
the area around the tip of the cassette is visibly
soiled,
the error message E-4 with the text “See User’s
Manual” is displayed; the error message might be
caused by uff, crumbs, etc. inside the tip of the
cassette.
To enable you to clean the inside of the meter, you
have to open Clean in the Tools menu and then remove
the test cassette (see next page).
i
If the error message E-4 was displayed, a test
area is lost when you later reinsert the test
cassette.

Table of Contents

Other manuals for Accu-Chek Mobile

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Mobile and is the answer not in the manual?

Accu-Chek Mobile Specifications

General IconGeneral
Measurement TechnologyPhotometric
Sample TypeCapillary whole blood
Sample Size0.3 µL
Test Time5 seconds
Battery Type2 AAA batteries
ConnectivityUSB
Dimensions121 x 63 x 20 mm
Weight129 g (with batteries)
DisplayLCD
Memory Capacity2000 results
Measuring Range0.6 - 33.3 mmol/L
Operating Temperature10 - 40°C
Operating Humidity10 to 90% RH
Hematocrit Range20 to 70%
Test Strip TechnologyNo test strips required (uses a test cassette)