218
Testing and storage conditions15
Do not use blood glucose results obtained
following these warnings as a basis for mak-
ing therapeutic decisions. These results may
be incorrect. Incorrect results can cause the
wrong therapy recommendation to be made
and thus produce serious adverse health
effects.
Tests cannot be performed at temperatures below
+8 °C and above +42 °C. In this event, the following
error message appears:
Never try to speed up the cooling or warming
of your meter, e.g. by placing it in the refriger-
ator or on a radiator. Doing so can damage
the meter and cause it to deliver incorrect
results. Incorrect results can cause the wrong
therapy recommendation to be made and
thus produce serious adverse health effects.
Do not store the lancing device in a primed state and
do not store it for any length of time in a place
where the temperature is very low or high, e.g. in a
hot car. This can cause parts of the lancing device to
become warped, so impairing its operation.

Table of Contents

Other manuals for Accu-Chek Mobile

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Mobile and is the answer not in the manual?

Accu-Chek Mobile Specifications

General IconGeneral
Measurement TechnologyPhotometric
Sample TypeCapillary whole blood
Sample Size0.3 µL
Test Time5 seconds
Battery Type2 AAA batteries
ConnectivityUSB
Dimensions121 x 63 x 20 mm
Weight129 g (with batteries)
DisplayLCD
Memory Capacity2000 results
Measuring Range0.6 - 33.3 mmol/L
Operating Temperature10 - 40°C
Operating Humidity10 to 90% RH
Hematocrit Range20 to 70%
Test Strip TechnologyNo test strips required (uses a test cassette)