Symbols, messages, problems and error messages16
E‑4
Fluff, crumbs, etc. are inside the tip of the cassette.
Remove these. To do this, follow the instructions in
Chapter 11.1.
i
You can go directly from here to the Clean
menu without having to turn the meter off
rst. To do this, press the
button (Con-
tinue).
The cassette is not properly inserted into the meter.
Remove the cassette and reinsert it.
The test area has absorbed too little blood or control
solution.
Repeat the test with a larger amount of blood or
control solution.
You did not apply the blood drop or the control
solution to the center of the test area.
Repeat the test. Apply the blood drop or the control
solution exactly to the center of the test area at the
tip of the cassette.

Table of Contents

Other manuals for Accu-Chek Mobile

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek Mobile and is the answer not in the manual?

Accu-Chek Mobile Specifications

General IconGeneral
Measurement TechnologyPhotometric
Sample TypeCapillary whole blood
Sample Size0.3 µL
Test Time5 seconds
Battery Type2 AAA batteries
ConnectivityUSB
Dimensions121 x 63 x 20 mm
Weight129 g (with batteries)
DisplayLCD
Memory Capacity2000 results
Measuring Range0.6 - 33.3 mmol/L
Operating Temperature10 - 40°C
Operating Humidity10 to 90% RH
Hematocrit Range20 to 70%
Test Strip TechnologyNo test strips required (uses a test cassette)

Related product manuals